6 conditioning / scaling, 4) measuring conditions, 5) operating range – VEGA VEGAPULS 56V User Manual

Page 64

Advertising
background image

64

VEGAPULS 56V

Signal
condit
ioning

Sca-
ling

0 %
corre-
sponds

XXXX

100 %
corre-
sponds

XXXX

Deci-
mal
point

888.8

prop.
to

Mass

Unit

Kg

Adjustment with medium

Example:
Min/Max-adjustment: 0,500 … 3,400 m; oper-
ating range to approx. 1,800 … 2,900 m.
Then the sensor indicates only measured
values from 20!…!60!%.

6 Conditioning / Scaling

Under the menu point "

Conditioning/Scaling“

you enter a figure corresponding to a 0 %
and 100 %-filling and choose beside the
position of the decimal point a physical unit,
e.g. distance (menu plan, page 66).

Enter in the menu window "

0 % corresponds“

the figure of the 0 %-filling (in the example of
the adjustment with PC this had been 45
liters).

• Confirm with "OK“.

With the "–>“-key you move to the 100!%
menu. Enter the figure of your parameter
corresponding to a 100 %-filling. In the exam-
ple this had been 1200 for 1200 liters.

• Confirm with "OK“.

Choose, if required, a decimal point. How-
ever note that only max. 4 digits can be
shown.

In the menu "

Prop. to“ you choose the param-

eter (mass, volume, distance…) and in the
menu "

Unit“ the physical unit (kg, l, ft

3

, gal, m

3

…).

Set-up

Fill the vessel e.g. to 10 % and enter in the
menu "

Min-adjustment“ with the "+“ and "–“-

keys 10 %. Then fill the vessel e.g. to 80 % or
100 % and enter in the menu "

Max-adjust-

ment“ with the "+“ and "–“-keys 80 % or
100 %. When you do not know the distance,
you have to sound.

Note:
The sensor can only provide levels within the
defined operating range. For detection of
levels outside the operating range, this range
must be corrected appropriately in the menu
"

Sensor optimisation/Meas. environment“

(under figure 2).

4) Measuring conditions

Enter here if you want to measure liquids or
solids and confirm the concerned meas.
conditions (menu plan page 72).

5) Operating range

Without special adjustment, the operating
range corresponds to the measuring range
in the menu "

Adjustment“. The measuring

range was already adjusted with the min./
max. adjustment. If a smaller operating range
is selected than previously adjusted in "

Ad-

justment“, the sensor is caused to provide
the measured values only in a limited range,
e.g. instead of 0!…!100!% in the range of
20!…!60!% (meas. window).

with
medium

Min-
adjust
at %

XXX.X

Max-
adjust
at %

XXX.X