Luminex IS Software Manual for Version 2.3

x

MAP Technology

24

PN 89-00002-00-110 Rev. B

high or low end. Examples of out-of-range sample labeling for non-
competitive assays are shown in Table 6. Figure 18 shows what a
standard curve for a non-competitive assay should look like.

Figure 18 Non-Competitive Assay

Examples of out-of-range sample labeling for competitive assays are
shown in Table 7. Figure 19 shows what a standard curve for a
competitive assay should look like.

Table 6. Non-Competitive Out-Of-Range Labeling

Condition

Concentration Label

MFI of Standard Referenced

Left of Curve

< min concentration standard

Lowest

Right of Curve

> max concentration standard

Highest

Table 7. Competitive Out-Of-Range Labeling

Condition

Concentration Label

MFI of Standard Referenced

Left of Curve

< min concentration standard

Highest

Right of Curve

> max concentration standard

Lowest