3 important safety advice – Multichannel Systems HiClamp Manual User Manual

Page 10

Advertising
background image

HiClamp Manual

4

2.3 Important Safety Advice

Warning: Make sure to read the following advice prior to installations of the HiClamp. If you do
not fulfill all requirements stated below, this may lead to malfunctions, breakage, or even fatal
injuries. Obey always the rules of local regulations and laws. Only qualified personnel should be
allowed to perform laboratory work. Work according to good laboratory practice to obtain best
results and to minimize risks.

The product has been built to the state of the art and in accordance with recognized safety
engineering rules. The device may only

 be used for its intended purpose;

 be used when in a perfect condition.

Improper use could lead to serious, even fatal injuries to the user or third parties and damage
to the device itself or other material damage.

Malfunctions which could impair safety should be rectified immediately.

2.3.1 High

Voltage

Electrical cords must be properly laid and installed. The length and quality of the cords must
be in accordance with local provisions.

Only qualified technicians may work on the electrical system. It is essential that the accident
prevention regulations and those of the employers' liability associations are observed.

 Each time before starting up, make sure that the mains supply agrees with the specifications

of the product.

 Check the power cord for damage each time the site is changed. Damaged power cords should

be replaced immediately and may never be reused.

 Check the leads for damage. Damaged leads should be replaced immediately and may never

be reused.

 Liquids may cause short circuits or other damage. Keep the power supply and the power cords

always dry. Do not handle it with wet hands.

2.3.2 Requirements for the installation

 The HiClamp weighs more than 47 kg. Always grip it tightly and do not carry it alone,

but with the aid of another person.

 The movements of the well plate carrier can lead to vibrations of the workbench on which

the HiClamp is set up. Therefore, the HiClamp must be set up on a rigid, vibration-free base.
The base must also be sufficiently solid to carry the weight of the device.

 The HiClamp should be operated only in an air conditioned room. A room temperature

of 20 °C (or less) is recommended. Make sure that the device is not subject to direct sunlight.
It may overheat.

 If the air cannot circulate freely around the external power supply, the device may overheat.

Do not shield the power supply by laying anything on top of the acryl glass cover.

 The external power supply is only for use with the HiClamp. Do not connect it to any other

instrument.

Advertising