TEVC Recording Protocol

33

Loop including a variable component

The loop including a variable element can be used for defining voltage step protocols. Whenever
you include the place holder "x" instead of a number, for example in the "Voltage Command"
window and you loop back to this definition, you can define in the appearing "LoopX" window
the values replacing the "x" incrementing with every loop count. In the shown example the
potential steps go from -100 mV to +20 mV in 20 mV increments. The holding potential between
the steps is -60 mV.

Next DNA

Adding this icon as the last step of a program sequence allows measurements on multiple
messages expressed in oocytes. A green line is automatically drawn indicating where the program
will loop back upon execution of this step.

End of the Sequence

This icon is normally placed at the end of the program sequence. Its execution causes the machine
to stop, close the perfusion and remove the last cell that was measured.