ZOLL AED Pro Rev J User Manual

Page 112

Advertising
background image

C

ONFIGURABLE

S

ETTINGS

C–4 www.zoll.com

9650-0350-01 Rev. J

Breathing Prompt Delay

Sets the period over which the unit issues the check
breathing prompts.

2 seconds

4 seconds

6 seconds — default

9 seconds

15 seconds

20 seconds

Responsiveness/Patient

Selects the wording for this prompt.

OFF — default

Check Responsiveness

Check Patient

Responsiveness/Patient Prompt Delay

Sets the duration before the next prompt or disables
this prompt.

2 seconds

4 seconds — default

6 seconds

8 seconds

Circulation/Pulse

Selects the wording for this and related prompts.

OFF — default

Check Circulation

Check Pulse

Circulation/Pulse Prompt Delay

Sets the duration before the next prompt or disables
this prompt.

2 seconds

5 seconds

7 seconds

10 seconds — default

15 seconds

20 seconds

Press Treatment/Shock

Selects the wording for this and related prompts.

Press Flashing Treatment Button

Press Flashing Shock Button
default

CPR

CPR Monitoring

When this option is checked (On) and CPR-D•padz or
CPR stat•padz are attached, the unit performs CPR
monitoring and prompts the rescuer.

On — default

Off

No-Shock CPR Period

Sets the duration of the CPR period following a
no-shock result for the first analysis in a stack.

When you select Extended, the device remains in the
CPR state indefinitely and displays the ANALYZE
softkey. You must press the ANALYZE softkey to move
to analysis.

30 seconds

60 seconds

90 seconds

120 seconds — default

150 seconds

180 seconds

Extended

Cycle through CPR

When this option is checked (On) and pads are
detached from the patient for more than one and
one-half minutes, the unit prompts the rescuer to
perform CPR.

On — default

Off

Configurable Option

Possible Values