ZOLL AED Pro Rev J User Manual

Page 88

Advertising
background image

C

HAPTER

7

T

ROUBLESHOOTING

AND

M

AINTENANCE

7–8

www.zoll.com

9650-0350-01 Rev. J

For instructions on placing the unit back into service, refer to “Preparing the Unit for Clinical
Use” on page 2-5.

10

Activate the simulator’s CPR function.

11

Verify that the metronome begins to beep.

Verify that the AED Pro unit issues the following voice and text prompts within 60 seconds
(unless CPR monitoring is disabled on the unit):

PUSH HARDER
GOOD COMPRESSIONS

Verify that the chest compression gauge functions properly.

12

After approximately 2 minutes of CPR, verify that the unit issues the voice and text prompt
STOP CPR.

13

Verify that the AED Pro unit begins a new ECG analysis.

14

Turn off the AED Pro unit and the simulator.

15

Verify that the Ready indicator displays a green check before disconnecting the simulator
and attaching defibrillation electrode pads.

Step

Action