Monitoring with Defibrillation Electrodes

With defibrillation electrodes in use, you can change from semiautomatic mode to ECG
monitoring mode.

If the unit detects a shockable rhythm while in ECG monitoring mode and defibrillation
electrodes are in use, the unit prompts you to check the patient and then changes to
semiautomatic mode.

For more information, refer to Chapter 3, “Semiautomatic Mode”.

Before you begin

Attach defibrillation electrode pads to the patient (refer to Chapter 3.)

Procedure

To change from semiautomatic to ECG monitoring mode:

Step

Action

Result

1

Press and hold the left (unlabeled) softkey
for at least 5 seconds.

The unit displays MONITOR as the current
mode, and the SEMI-AUTO softkey label
appears.

2

Monitor the patient’s ECG.

If it detects a shockable rhythm while
monitoring, the unit prompts you to check the
patient and changes to semiautomatic mode.

3

To manually return to semiautomatic mode,
press and release the SEMI-AUTO softkey.

00:24:18

1

34

MONITOR

SEMI-AUTO

ECG X1.0

CHECK PATIENT