Descriptions of AED Pro Configurable Settings

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

C–7

Start CPR

At the beginning of the first CPR period, if Start
with CPR Period is enabled, the AED Pro issues
one of the voice and text prompts shown in the
Possible Values column.

Open Airway, Check Breathing,
Check Pulse, If no Pulse Start
CPR

Open Airway, Check Breathing,
Check Circulation, If No
Circulation Start CPR

Start CPR — default

Start Compressions

If no Pulse, Start Compressions

If no Circulation, Start
Compressions

Continue CPR

When you select one of the prompts in the
Possible Values column, that prompt will be
repeated once every 10 seconds (default) during
the CPR period if compression stops while in the
CPR state.

When this option is set to Off, no prompts will be
issued.

OFF — default

If no Circulation, Continue CPR

If no Pulse, Continue CPR

Continue CPR

Continue Compressions

If no Pulse, Continue
Compressions

If no Circulation, Continue
Compressions

Delays

Continue CPR Delay

Allows you to set the inverval between Continue
CPR
prompts.

5 seconds

10 seconds — default

15 seconds

Push Harder Delay

Allows you to set the inverval between Push Harder
prompts.

15 seconds — default

20 seconds

30 seconds

Audio and Text Prompts

Start CPR

Allows you to select whether you want to both hear and
view the START CPR prompt , or view it only.

Audio and Text — default

Text only

Push Harder

Allows you to select whether you want to both hear and
view the Push Harder prompt , or view it only.

Audio and Text — default

Text only

OFF

Good Compressions

Allows you to select whether you want to both hear and
view the Good Compressions prompt , or view it only.

Audio and Text — default

Text only

OFF

Continue CPR

Allows you to select whether you want to both hear and
view the Continue CPR prompt , or view it only.

Audio and Text — default

Text only

Stop CPR

Allows you to select whether you want to both hear and
view the Stop CPR prompt , or view it only.

Audio and Text — default

Text only