Descriptions of AED Pro Configurable Settings

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

C–5

Post-Shock CPR Period

Sets the duration of the CPR period following the
delivery of one or more shocks.

When you select Extended, the device remains in the
CPR state indefinitely and displays the ANALYZE
softkey. You must press the ANALYZE softkey to move
to analysis.

30 seconds

60 seconds

90 seconds

120 seconds — default

150 seconds

180 seconds

Extended

Start with CPR (Semiautomatic Mode only)

When this option is checked (On), before beginning
ECG analysis, the unit prompts the rescuer to check the
patient’s pulse and perform a period of CPR if no pulse
is detected.

On

Off — default

Start with CPR Period

Sets the duration of the CPR period associated with the
Start with CPR option.

When you select Extended, the device remains in the
CPR state indefinately and displays the ANALYZE
softkey. You must press the ANALYZE softkey to move
to analysis

30 seconds

60 seconds

90 seconds

120 seconds — default

150 seconds

180 seconds

Extended

Analyze Key

Enables or disables the ANALYZE softkey during the
Post Shock and Post No Shock CPR periods. If the
ANALYZE softkey is enabled, it appears during both
the Post Shock and Post No Shock CPR periods.

On

Off — default

Fully Release

When checked (On), this text prompt reminds the
rescuer to lift the hands off the chest during CPR to
allow full recoil.

On

Off — default

CPR Prompts

Start with CPR Period

Start CPR

At the beginning of the first CPR period, if Start
with CPR Period is enabled, the AED Pro issues
one of the voice and text prompts shown in the
Possible Values column.

Open Airway, Check Breathing,
Check Pulse, If no Pulse Start
CPR

Open Airway, Check Breathing,
Check Circulation, If No
Circulation Start CPR

Start CPR — default

Start Compressions

If no Pulse, Start Compressions

If no Circulation, Start
Compressions

Configurable Option

Possible Values