C

HAPTER

3

S

EMIAUTOMATIC

M

ODE

3–4

www.zoll.com

9650-0350-01 Rev. J

4

Hold the CPR sensor in place with your left hand, and
use your right hand to pull the number 3 tab and peel
the protective backing from the electrode.

Press the electrode from the center of the abdomen
outward to push out air from beneath the pad, and to
make sure the pad adheres properly to the patient’s
skin.

If the patient is large or there is a need to place the
electrode under a breast, you can detach the lower pad
at the perforation and extend it for effective positioning.

Place the pad slightly to the patient’s left and below the
patient’s left breast.