Start with cpr option, Messages in semiautomatic mode – ZOLL AED Pro Rev J User Manual

Page 52

Advertising
background image

C

HAPTER

3

S

EMIAUTOMATIC

M

ODE

3–12

www.zoll.com

9650-0350-01 Rev. J

Start with CPR Option

Your AED Pro unit may be configured to start a rescue with a CPR period of 30 to 180 seconds
that begins after you power on the AED Pro unit. You can end this initial CPR period at any
time by pressing the Analyze softkey. When you press the Analyze softkey, the AED Pro unit
immediately begins analyzing the patient’s ECG rhythm if the electrode pads are attached
correctly. The Analyze softkey appears only during the initial CPR period and does not appear
during any subsequent CPR periods unless it is configured to do so and/or the subsequent CPR
intervals are set to Extended.

Messages in Semiautomatic Mode

During semiautomatic mode, the unit can issue the following voice prompts and text messages.
The unit issues each voice prompt only once, but the equivalent message remains on the screen
until you take action, time expires, or the device status changes.

Message

Description

UNIT OK

The unit successfully passed the power-on self-test.

UNIT FAILED

The unit failed the power-on self-test and is not usable for
patient care.

CHANGE BATTERY

The self-test detected a low energy condition that is insufficient
for patient care. Replace the battery pack immediately.

ADULT PADS

PEDIATRIC PADS

The unit detected the specified type of electrode pads and
adjusted defibrillation energy settings accordingly.

PLUG IN CABLE

The unit started up without an electrode cable plugged in. Plug
the cable into the unit.

ATTACH DEFIB PADS TO
PATIENT’S BARE CHEST

The unit does not detect attachment of the defibrillation
electrode pads to the patient.

Check the cable for damage.

Make sure that the pads are properly applied to the patient.

If this message continues, check the electrode sites to ensure
that they are clean, dry, and free of excess hair. Check the
expiration date on the electrode package.

CHECK PATIENT

Check the patient for responsiveness or consciousness by
gently shaking the patient and shouting “Are you all right?”

CHECK PULSE

Check the patient for a pulse or other signs of circulation, such
as normal breathing, movement, or coughing.

START CPR

Indicates the beginning of a CPR period. Begin CPR.

STOP CPR

Indicates the end of the CPR period. Discontinue CPR.

DON’T TOUCH PATIENT,
ANALYZING

Indicates the beginning of an ECG analysis period. Make sure
everyone stands clear of the patient. Keep the patient
motionless during ECG analysis.

NO SHOCK ADVISED

ECG analysis did not detect a shockable rhythm.

SHOCK ADVISED

ECG analysis detected a shockable rhythm that requires
therapy.

Advertising