Guidance and Manufacturer’s Declaration — Electromagnetic Emissions

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

A–7

Electromagnetic Immunity Declaration (EID)

The ZOLL AED Pro device is intended for use in the electromagnetic environment specified
below. The customer or operator should ensure that the device is used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance

Level

Electromagnetic Environment –

Guidance

Electrostatic
discharge
(ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.

Electrical fast
transient/burst

IEC 61000-4-4

±2 kV for power supply
lines

±1 kV for input/output
lines

Not applicable

±1 kV I/O

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV for common
mode

Not applicable

Not applicable

Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines

IEC 61000-4-11

<5% Ut (>95% dip in
Ut) for 0.5 cycle

40% Ut (60% dip in Ut)
for 5 cycles

70% Ut (30% dip in Ut)
for 25 cycles

<5% Ut (>95% dip in
Ut) for 5 seconds

Note: Ut is the ac
mains voltage prior to
application of the test
level.

Not applicable

Not applicable

Not applicable

Not applicable

Power
frequency
(50/60 Hz)
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.