ZOLL AED Pro Rev J User Manual

Page 114

Advertising
background image

C

ONFIGURABLE

S

ETTINGS

C–6 www.zoll.com

9650-0350-01 Rev. J

Continue CPR

When you select one of the prompts in the
Possible Values column, that prompt will be
repeated once every 10 seconds (default) during
the CPR period if compression stops while in the
CPR state

When this option is set to Off, no prompts will be
issued.

OFF — default

If no Circulation, Continue CPR

If no Pulse, Continue CPR

Continue CPR

Continue Compressions

If no Pulse, Continue
Compressions

If no Circulation, Continue
Compressions

Post Shock CPR Period

Start CPR

At the beginning of the CPR period, after the
ECG analysis, the AED Pro issues one of the
voice and text prompts shown in the Possible
Values column.

Open Airway, Check Breathing,
Check Pulse, If no Pulse Start
CPR

Open Airway, Check Breathing,
Check Circulation, If No
Circulation Start CPR

Start CPR — default

Start Compressions

If no Pulse, Start Compressions

If no Circulation, Start
Compressions

Continue CPR

When you select one of the prompts in the
Possible Values column, that prompt will be
repeated once every 10 seconds (default) during
the CPR period if compression stops while in the
CPR state.

When this option is set to Off, no prompts will be
issued.

OFF — default

If no Circulation, Continue CPR

If no Pulse, Continue CPR

Continue CPR

Continue Compressions

If no Pulse, Continue
Compressions

If no Circulation, Continue
Compressions

Post No Shock CPR Period

Start CPR

At the beginning of the CPR period after the
ECG analysis, the AED Pro issues one of the
voice and text prompts shown in the Possible
Values column.

Open Airway, Check Breathing,
Check Pulse, If no Pulse Start
CPR

Open Airway, Check Breathing,
Check Circulation, If No
Circulation Start CPR

Start CPR — default

Start Compressions

If no Pulse, Start Compressions

If no Circulation, Start
Compressions

Configurable Option

Possible Values