Important: If you change the battery while the unit is in clinical use (that is, with a cable

connected), the unit automatically powers on in nonrescue mode and then shuts
down. Press and release the On/Off button to restart the unit.

CHANGE BATTERY Warning

When the unit detects a low energy condition, it issues the voice and text prompt CHANGE
BATTERY
once every minute. Depending on the age and condition of the battery, the remaining
operating time of the unit might be extremely limited.

The warning message continues until the device shuts down.

WARNING!

To ensure the availability of adequate power during an emergency, keep a fully charged
spare battery pack with the device at all times.

Whenever the unit issues the prompt CHANGE BATTERY, immediately replace the used
battery pack with a fully charged one to ensure continuous operation and to avoid unexpected
device shutdown. After removing a depleted rechargeable battery pack from the unit, recharge
the pack as soon as possible.

Battery Condition

Indications

Correction

Low energy detected during
power-on self-test.

Message:
CHANGE BATTERY

Replace battery pack.

Low energy or other self-test
failure while the unit is
powered off (standby).

Ready indicator shows a
red “X”. Unit beeps once every
minute for 30 minutes.

Replace battery pack.

Check or replace
preconnected electrodes.

If the red “X” remains, contact
ZOLL Technical Service.

Low energy detected while the
unit is powered on.

Message:
CHANGE BATTERY

Replace battery pack as soon
as possible.

Dead battery

Ready indicator shows a
red “X”.

Replace battery pack.

If the red “X” remains, contact
ZOLL Technical Service.