Messages
While preparing the AED Pro unit for use, the following messages can be seen and/or heard:
Message
Description
UNIT OK
The unit successfully passed its power-on self-test.
UNIT FAILED
The unit failed its power-on self-test and is not usable
for patient care.
CHANGE BATTERY
The self-test detected a low energy condition that is
insufficient for patient care. Replace the battery pack
immediately.
ADULT PADS
PEDIATRIC PADS
The unit detected the specified type of electrode pads
and adjusted defibrillation energy settings accordingly.
PLUG IN CABLE
The unit started up without an electrode cable plugged
in. Plug the cable into the unit.
NON-RESCUE MODE
The device is operating in nonrescue mode and an IrDA
connection is established.
POWERING OFF
The On/Off button was pressed and held for 1 second
to turn the unit off.