Before you begin

Be sure the defibrillation cable is plugged into the unit.

Procedure

To defibrillate the patient in semiautomatic mode:

Step

Action

Result

1

Press and release the On/Off button to turn
on the unit.

On successful completion of the power-on
self-test, the unit indicates the type of
attached electrodes. If the defibrillation
pads are not attached to the patient, the
unit issues the voice and text prompt:

ATTACH DEFIB PADS TO PATIENT’S

BARE CHEST

2

If prompted, apply defibrillation electrode
pads to the patient.

(Refer to “Applying Defibrillation Electrode
Pads” on page 3-2.)

When the pads are properly attached to the
patient, the unit issues the voice and text
message:

DON’T TOUCH PATIENT, ANALYZING

and then begins ECG analysis.

3

Allow the unit to analyze the patient’s ECG.

The unit determines whether or not the
patient has a shockable rhythm and then
displays and voices its recommendation:

SHOCK ADVISED

or

NO SHOCK ADVISED

4

After ECG analysis, which message
appears?

— If SHOCK ADVISED, continue with step 5.

— If NO SHOCK ADVISED, go to step 8.

If a shock is needed, the defibrillator
automatically begins charging.

If a shock is not needed, the unit prompts
you to perform CPR.

5

Wait for the defibrillator to charge.

When fully charged, the unit emits a
charge-ready tone, repeatedly flashes the
Shock button, and issues the voice and
text prompts:

DON’T TOUCH PATIENT

PRESS FLASHING SHOCK BUTTON