ZOLL Propaq MD Rev B User Manual

Page 24

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

20

9650-0801-01 Rev. B

6.5

On the Propaq MD unit, do the
following:

Press the More (

) then the

Setup (

) quick access keys.

Use the navigation keys to
select Supervisor>Display/
Configuration>Device Info
and select Restore All Factory
Defaults
.

Note: You need a supervisor
passcode to enter the Supervisor
menu.

6.6

Press the “I Measure” switch on the
Battery Current Test Fixture.

Observe the current measurement on
the DMM, and verify that the reading
is between 0.47mA-1.25mA.

o

o

Do this...

Observe this...

Pass/Fail

Advertising