10.0 Pacer Test

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent)

Do This...

Observe This...

Pass/Fail

10.1 On the Analyzer, select the PACER

function per test equipment instructions.

10.2 On the Analyzer, press F2 to select PACER

JACKS, then press F5 (Done) to measure
pulse.

10.3 Connect the MFC connector to the

Analyzer.

10.4 On the Propaq MD unit, apply power to the

unit using a battery or an AC adapter.

10.5 On the Propaq MD unit, press the PACER

button.

10.6 On the Propaq MD unit, use the navigation

keys to do the following:

•Set Rate to 30.

•Set Mode to Fixed.

•Set Output to 10.

•Select Start Pacer.

o

o

10.7 On the Impulse 7000 DP Analyzer, record

the following:

•pulse rate

•pacer amplitude output

•pacer pulse width

•Pacer Rate PPM
(28.5ppm-31.5ppm)

•Pacer amplitude output
(5mA-15mA)

•Pacer pulse width (38ms to 42ms)

Note: If an ECG FAULT message
displays when leads are not
attached, it is normal.

10.8 On the Propaq MD unit, use the navigation

keys to do the following:

•Set Rate to 180.

•Set Output to 140.

o

o

10.9 On the Impulse 7000 DP Analyzer, record

the following:

•pulse rate

•pacer amplitude output

•pacer pulse width

•Pacer Rate PPM (162ppm-198ppm)

•Pacer amplitude output
(133mA-147mA)

•Pacer pulse width (38ms to 42ms)

10.10 Select Turn Pacer Off.