HAPTER

6

F

UNCTIONAL

D

ESCRIPTION

86

9650-0801-01 Rev B

module contains power supplies, charging and discharge circuitry, pacing circuitry, and the
Pads/Paddles ECG preamplifier and impedance circuitry. Communication with the
Defibrillator/Pacer module is a function of the PP. The ECG-Charger board contains the
charger circuit, ECG acquisition circuitry, and protection components. The
Processor-Bridge-Pacer board contains the micro controller, CPLD, Pacer, and defibrillation
shock delivery circuit.

Defib CPLD

The Defib CPLD provides I/O control for the Pacer circuitry, the defibrillator charging and fire
control circuitry.

Pace/Defib (PD) Micro controller

The PD micro controller communicates with the PP processor on the monitor board to accept
commands to control the defibrillator and pacer. The PD micro controller controls the pacer
current and pulse timing and the defibrillation charging and waveform generation.
Communication between the Monitor and Defibrillator/Pacer boards is routed through the flex
cable connected to J106 of the defibrillator board and J38 on the Monitor board.

Charging Circuit

This is a high voltage step-up inverter that charges the Defib Capacitor. It provides electrical
isolation between the defib output and the system. The charging circuit will charge the Defib
Capacitor to up to 2775V. The Defib Capacitor is connected to the Defibrillator/Pacer board
through the 2-wire cable connected to J119.

Fire and Relay Circuit

The fire circuit employs high voltage switching devices to deliver a biphasic waveform to the
patient. Switching devices are utilized to produce a Rectilinear Biphasic™ waveform. In
addition, relays are located between the fire circuit and the patient for electrical isolation. They
are engaged only when a valid command is set from the Defib CPLD.

Pacer

Pacer circuit delivers pacing pulses under control of the PD micro controller over an isolation
transformer. The pacer circuitry consists of an adjustable constant current power supply, pulse
timing, and current monitoring circuits.

Pads/Paddles ECG Preamplifier

This is a single channel ECG preamplifier paddle pick up that monitors the patient ECG signal
from the Defib paddles (or pads). It digitizes this signal and delivers it to the PP processor via
the PP interface. This preamp is a simplified version of the 3/5/12 Lead ECG utilized on the
Monitor Board.

Paddle Interface

Paddle interface circuitry detects analog voltage levels to decode switches and ID resistors of
the standard ZOLL paddles and internal paddles. The paddle set interface is connected to J115
of the Processor-Bridge-Pace PCB.