HAPTER

2

T

ROUBLESHOOTING

52

9650-0801-01 Rev B

CO2 GAS INPUT
OCCLUSION

CO

2

inlet blocked.

Replace FilterLine

®

. If

problem persists, check for
CO

2

inlet occlusion. If still

persists, replace the CO

2

module.

CO2 MAINTENANCE
REMINDER

Periodic maintenance
reminder.

Replace CO

2

module soon.

CO2 MAINTENANCE
REQUIRED

Periodic Maintenance due.

Replace CO

2

module.

CO2 NO FILTERLINE

No filterline is detected.

If FilterLine is connected
and problem persists,
replace FilterLine. If still
persists, replace the CO

2

module.

CO2 CAL FAILURE

Calibration failed.

Retry calibration or
continue with old
calibration. Replace the
CO

2

module.

DEFIB PACER DEVICE
FAILURE

One of the following issues
occurred:

Error was detected on the
defib/pacer board.

Interface self-test
communication failed.

Pacer interface self-test
failed.

Communication with
pacer hardware failed
after startup.

Check communication
cables. If problem persists,
replace the defib board
module.

DEFIB DEVICE FAILURE

One of the following issues
occurred:

Device could not charge
up the self-test energy.

Discharge circuit failed to
discharge voltage within
specified time duration.

Device failed to deliver
correct self-test energy.

Check communication
cables. If problem persists,
replace defib board
module.

FP DEFIB FIRE SHORT

Front panel button is stuck.

Replace the front panel.

PS DEFIB CHARGE
SHORT

Paddle set fire button is
stuck or paddle set energy
up select button is stuck.

Replace the paddles.