Propaq MD Overview

9650-0801-01 Rev. B

Propag MD Service Manual

29

12.0 Treatment Report Test

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent)

Do this...

Observe this...

Pass/Fail

12.1 Connect the multifunction cable (MFC) to the

Analyzer, and turn on the Analyzer.

12.2 Turn on the Propaq MD unit, and press the

following quick access keys:

a)

b)

c)

The Clear Log window
displays.

12.3 Select Yes.

The unit displays a clearing
the log message.

o

o

12.4 Press the CHARGE button, select 200 joules,

then press the CHARGE button again.

The unit starts to charge, a
charging tone sounds, and the
Shock button illuminates.

o

o

12.5 Press the Shock button.

The unit discharges the shock.

o

o

12.6 Wait 20 seconds, press the Treatment quick

access key(

), and select the first code

marker.

12.7 Wait 15 seconds and turn off the unit, then wait

2 minutes and turn on the unit.

12.8 On the Propaq MD unit, press the following

quick access keys:

a)

b)

c)

The Treatment Summary
window displays.

o

o

12.9 Highlight the second event and press Select.

Verify that a green check mark
displays next to the file.

o

o

12.10 Select Print Treatment Summary.

Verify that the recorder prints
out the defib event and the
code marker snapshot.

o

o

12.11 Turn off the Propaq MD unit.