Propaq MD Error Messages

9650-0801-01 Rev B

ZOLL Propaq MD Service Manual

53

PS DEFIB ENERGY DOWN
SHORT

Paddle set energy down
select button is stuck.

Replace the paddles.

PS DEFIB RECORDER
SHORT

Paddle set recorder button is
stuck.

Replace the paddles.

DEFIB CAP LEAKAGE

Capacitor bank cannot hold
a charge.

Replace the capbank
module.

DEFIB USING DEFAULT
CALIBRATION

Calibration data lost.

Replace the defib board
module.

DEFIB FATAL CHARGE
ERROR

Charging error.

Replace the defib board
module.

DEFIB CHARGE ERROR

Over voltage was detected.

If problem persists, replace
the defib board module.

DEFIB LOW BATTERY

Battery is too low to charge
defibrillator.

Charge the battery. Check
battery contacts and clean
if necessary.

DEFIB CHARGE TIMEOUT
FAIL

The battery did not produce
enough current to charge
capacitor.

Charge the battery. If
voltages are okay (from
battery and aux), replace
the defib board module.

DEFIB DISCHARGE TEMP

Device is in a hot
environment or the
discharge circuit is always
on.

Move device to cooler area.
If problem persists, replace
the defib board module.

DEFIB FIRE FAILURE

High impedance or low
impedance (short).

Replace pads. If problem
persists, check internal
connections. If still persists,
replace the defib board
module.

DEFIB PHASE FAILURE

Phase I or phase II
waveform dropout.

Replace the defib board
module.

DEFIB NO ENERGY
DELIVERED

High impedance. No defib
energy was delivered.

Replace pads. If problem
persists, check internal
connections. If still persists,
replace the defib board
module.

PACER DEVICE FAILURE

Generation of pacer pulse
failed.

Replace the defib board
module.

PACER CURRENT FAULT

Lack of pacer current.

Replace the defib board
module.

Error Message

Explanation

Technical Action