ZOLL Propaq MD Rev B User Manual

Page 77

Advertising
background image

Removing the Defibrillator Board and Capacitor Bank Modules

9650-0801-01 Rev B

ZOLL Propaq MD Service Manual

73

Removing the Defibrillator Board and Capacitor Bank
Modules

Tools Required

#1 Phillips screwdriver with torque capability

#2 Phillips screwdriver with torque capability

26 mm open end wrench

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 57

2.

“Removing the Printer Module” on page 58

3.

“Disassembling the Main Unit” on page 60

Note:

After replacing the Defibrillator Board and/or Capacitor Bank modules, see Chapter 4,
"Resetting Life Cycle Counters" to
reset the life cycle counter for the module.

To reinstall the defibrillator board and/or capacitor bank modules, reverse
the steps.

Defibrillator Board and Capacitor Bank Disassembly

1

Lay back end of separated unit down with
opened side facing up.

2

Remove interface board and disconnect
the black cable by pulling the black snap
rivet straight out.

3

Remove the multi-color cable from the
defibrillator board.

4

Pull out small interconnect board.

multi-color cable

Advertising