0 ibp test, Ibp test – ZOLL Propaq MD Rev B User Manual

Page 38

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

34

9650-0801-01 Rev. B

17.0 IBP Test

Tools Needed

IBP Simulator (with accuracy better than +/- 1 mmHg), or equivalent

Setup

Connect the test cables to IBP patient simulator.

Do This...

Observe This...

Pass/Fail

17.1 Connect the IBP patient simulator to the

Propaq MD P1 channel.

17.2 Set the IBP patient simulator to 0 mmHg.

17.3 On the Propaq MD unit, press > IBP.

17.4 On the Propaq MD unit, Press > P1->0<-. Display of 0/0(+/-2) in the P1 box

at the bottom of the display (P1 is
zeroed).

The waveform for channel 1 is
present.

o

o

o

o

17.5 Exit the menu to see the waveform.

17.6 Set the IBP patient simulator to 200

mmHg (or as close to 200 mmHg as your
simulator allows).

Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).

o

o

17.7 Connect the IBP patient simulator to the

Propaq MD P2 channel.

17.8 Set the IBP patient simulator to 0 mmHg.

17.9 On the Propaq MD unit, press > IBP.

17.10 On the Propaq MD unit, Press > P2->0<-. Display of 0/0(+/-2) in the P2 box

at the bottom of the display (P1 is
zeroed).

The waveform for channel 2 is
present.

o

o

o

o

17.11 Exit the menu to see the waveform.

17.12 Set the IBP patient simulator to 200

mmHg (or as close to 200 mmHg as your
simulator allows).

Display of 200/200(+/-5) in the P2
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P2
diastolic pressure is 200 (+/-5)).

o

o

17.13 Connect the IBP patient simulator to the

Propaq MD P3 channel.

17.14 Set the IBP patient simulator to 0 mmHg.

17.15 On the Propaq MD unit, press > IBP.