17.16 On the Propaq MD unit, Press > P2->0<-. Display of 0/0(+/-2) in the P3 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 3 is
present.
o
o
o
o
17.17 Exit the menu to see the waveform.
17.18 Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
Display of 200/200(+/-5) in the P3
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P3
diastolic pressure is 200 (+/-5)).
o
o
17.19 Disconnect IBP cable.
Do This...
Observe This...
Pass/Fail