ZOLL Propaq MD Rev B User Manual

Page 39

Advertising
background image

Propaq MD Overview

9650-0801-01 Rev. B

Propag MD Service Manual

35

17.16 On the Propaq MD unit, Press > P2->0<-. Display of 0/0(+/-2) in the P3 box

at the bottom of the display (P1 is
zeroed).

The waveform for channel 3 is
present.

o

o

o

o

17.17 Exit the menu to see the waveform.

17.18 Set the IBP patient simulator to 200

mmHg (or as close to 200 mmHg as your
simulator allows).

Display of 200/200(+/-5) in the P3
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P3
diastolic pressure is 200 (+/-5)).

o

o

17.19 Disconnect IBP cable.

Do This...

Observe This...

Pass/Fail