Index – Yokogawa DAQWORX User Manual

Page 265

Advertising
background image

7-89

IM WX101-01E

1

2

3

4

5

6

7

8

9

Index

Setting the Recorder

Scan Interval

Scan interval

Select the scan interval. You cannot select fast sampling mode (125 ms) on the following

models:

• Models equipped with external input channels (/MC1 option)

• Models with the multi batch function (/BT2 option; release number 3 or later)

A/D integrate

Select the A/D integration time as necessary. Only the selectable settings are displayed.

Settings

Description

Auto

The DX automatically detects the power supply frequency and sets the integration
time to 16.7 ms and 20 ms for 60 Hz and 50 Hz, respectively. Fixed to 20 ms on
/P1 models that use the 24 VDC power supply.

50Hz

Sets the integration time to 20 ms.

60Hz

Sets the integration time to 16.7 ms.

100ms

Sets the integration time to 100 ms (when the scan interval is 2 s or 5 s).

600Hz

The A/D integration time for fast sampling mode. You cannot change this value.
You cannot use fast sampling mode on models with the external input channel
(/MC1) option. You cannot use fast sampling mode when the multi batch function
(/BT2 option) is being used

7.7 Setting DX1000/DX2000