Characteristics – Agilent Technologies 16700 User Manual

Page 101

Advertising
background image

101

16534A Specifications*

Bandwidth

dc to 500 MHz

dc offset accuracy

±(1% of offset + 2% of full scale)

dc voltage measurement accuracy

±(1.5% of full scale + offset accuracy)

Time interval measurement

±[(0.005% of D T) + (2E–6 x delay setting) + 100 ps]

accuracy at maximum sampling rate,
on a single scope card, on a single
acquisition

Trigger sensitivity (See notes)
• dc to 50 MHz

• 0.06 full scale

• 50 MHz to 500 MHz

• 0.13 full scale

Input resistance

1 M

Ω ±1% 50 Ω ±1%

* Specifications refer to the input to the BNC connector

Notes:

Specifications apply only within ± 10° C of the temperature at which the most recent calibration was performed.

Specifications apply only after operational accuracy calibration is performed in the frame in which the
oscilloscope module is installed.

Display magnification is used below 56 mV full scale. For sensitivities from 16 mV to 56 mV full scale, full scale is
defined as 56 mV.

Characteristics

General

Maximum sampling rate

2 GSa/s

Number of channels

• 2 to 8 using the same

time base and trigger.

• Up to 10 channels may be installed in a single

16700 frame, or up to 20 in a single system using a
16701 expansion frame.

Waveform record length

32768 points

Oscilloscope Modules Specifications and
Characteristics