Agilent Technologies 16700 User Manual

Page 103

Advertising
background image

103

Oscilloscope Modules Specifications and
Characteristics

16534A Characteristics

Horizontal (Time)

Time base ranges

0.5 ns/div to 5 s/div

Time base resolution

10 ps

Delay range
• pretrigger

• -32 K x sample period

• posttrigger

• 320 ms or 1.6E7 x sample period, whichever is greater

Time interval measurement accuracy

±{(0.005% of

∆ T) + (2E–6 x delay setting)

forsampling rates other than maximum,

+ [0.15/(sample rate)]}

for bandwidth-limited signals [signal
rise time > 1.4/(sampling rate)], on a
single card, on a single acquisition

Time interval measurement accuracy

± [(0.005% of

∆T) + (2E–6 x delay setting) + 300 ps

for 2, 3, or 4 Agilent 16533As or 16534As
operating on a single time base, for
measurements made between channels
on different cards, at maximum
sampling rate

Trigger

Trigger level range (See notes)

±1.5 x full scale from center of screen

Trigger modes
• Immediate

• Triggers immediately after arming condition is met

• Edge

• Triggers on rising or falling edge on channel 1 or

channel 2

• Pattern

• Triggers on entering or exiting a specified pattern

across both channels

• Auto condition

• Self-triggers if trigger is not satisfied within

approximately 50 ms after arming

• Events delay

• The trigger can be set to occur on the nth occurrence

of an edge or pattern, n

≤ 32000

• Intermodule

• Arms another measurement module or activates the

port out BNC connector when the trigger condition is
met

Notes:

Specifications apply only within ± 10° C of the temperature at which the most recent calibration was performed.

Specifications apply only after operational accuracy calibration is performed in the frame in which the
oscilloscope module is installed.

Display magnification is used below 56 mV full scale. For sensitivities from 16 mV to 56 mV full scale, full scale is
defined as 56 mV.