Agilent Technologies 16700 User Manual

Page 95

Advertising
background image

95

State/Timing Modules Specifications and
Characteristics

Agilent Technologies 16760A
Supplemental Specifications* and Characteristics

Probes

E5378A Single-ended

E5379A Differential

E5380A Mictor

E5382A Single-Ended Flying Leads

Input resistance and

Refer to figure 6.14

Refer to figure 6.14

Refer to figure 6.14

Refer to figure 6.15

capacitance

Maximum state data

1.5 Gb/s

1.5 Gb/s

600 Mb/s

1.5 Gb/s

rate supported

Mating connector

Agilent part number

Agilent part number

Amp Mictor 38 [2]

None required

1253-3620 [1]

1253-3620 [1]

Minimum voltage swing

250 mV p-p

Vin

+ - V

in

- >= 200 mV p-p

300 mV p-p

250 mV p-p

Input dynamic range

-3 Vdc to +5 Vdc

-3 Vdc to +5 Vdc

-3 Vdc to +5 Vdc

-3 Vdc to +5 Vdc

Threshold accuracy

+/- (30 mV + 1% of setting)*

+/- (30 mV + 1% of setting) [3]

+/- (30 mV + 1% of setting)

+/- (30 mV + 1% of setting)

Threshold range

-3.0 V to +5.0 V

-3.0 V to +5.0 V

-3.0 V to +5.0 V

-3.0 V to +5.0 V

User-supplied threshold

-3.0 V to +5.0 V

N/A

N/A

N/A

input range

User-supplied threshold

>= 100K ohms

N/A

N/A

N/A

input resistance

Threshold control options

• User-provided input

If operated single-ended

Adjustable from user

Adjustable from user

• Adjustable from user

(minus inputs grounded),

interface

interface

interface

the threshold can be adjusted
from the user interface

Maximum nondestructive +/-40 Vdc

+/-40 Vdc

+/-40 Vdc

+/-40 Vdc

input voltage

Maximum input slew rate

5 V/ns

5 V/ns

5 V/ns

5 V/ns

Clock input

Differential

Differential

Single-ended

Differential

Number of inputs [4]

34 (32 data and 2 clock/data) 17 (16 data and 1 clock/data)

34 (32 data and 2 clock/data) 17 (16 data and 1 clock/data)

* All specifications noted by an asterisk are the performance standards against which the product is tested.
[1] A support shroud, Agilent part number 16760-02302 (for boards up to 0.062" thick) or 16760-02303 (for boards up to 0.120" thick) is recommended.

A kit of 5 shrouds and 5 connectors is available as Agilent part number 16760-68702 (for boards up to 0.062" thick) or 16760-68703 (for boards up to 0.120" thick).

[2] A kit of 5 Amp Mictor connectors and 5 support shrouds is available, Agilent part number E5346-68701.

A support shroud is available separately, Agilent part number E5346-44701.

[3] If operated single-ended (minus inputs grounded), the threshold can be adjusted from the user interface.
[4] Refer to specifications on specific modes of operation for details on how inputs can be used.

R

1

R

2

C

1

0.7pF

20k

+0.75 V

Model Number

C

1

R

1

R

2

E5378A, E5379A

1.5pF

120

30

E5380A

3pF

120

60

Figure 6.14. E5378A, E5379A, E5380A input
equivalent probe load.

121

215

30

1pF

0.6pF

20k

+0.75 V

Figure 6.15. E5382A input equivalent probe load, with
5cm damped wire (see user’s guide for load models
with other accessories).