Pattern Generation Modules Specifications
and Characteristics

Agilent E8141A LVDS Data Pod

Output type

65LVDS389 (LVDS data lines)
10H125 (TTL non-3-state channel 7)

3-state enable

positive true TTL; no connect=enabled

Maximum clock

300 MHz

Skew

typical < 1 ns; worst case = 2 ns

Recommended lead set:

E8142A

Recommended lead set

Agilent 10498A

10 K

65LVDS389
ENABLE

°

3.3 V

LVDS DATA OUT

3-STATE IN TTL

Agilent 10476A 3-State 1.8 Volt Data Pod

Output type

74AVC16244

3-state enable

negative true, 38 K

Ω to GND, enabled on no connect

Maximum clock

300 MHz

Skew [1]

typical < 1.5 ns; worst case = 2 ns

Recommended lead set

Agilent 10498A

Agilent 10483A 3-State 3.3 Volt Data Pod

Output type

74AVC16244

3-state enable

negative true, 38 K

Ω to GND, enabled on no connect

Maximum clock

300 MHz

Skew [1]

typical < 1.5 ns; worst case = 2 ns

Recommended lead set

Agilent 10498A

74AVC16244

74AVC16244

[1] Typical skew measurements made at pod connector with approximately 10 pF/50 K

Ω load to GND; worst case

skew numbers are a calculation of worst case conditions through circuits. Both numbers apply to any channel
within a single or multiple module system.