Pattern Generation Modules Specifications
and Characteristics

[1] Typical skew measurements made at pod connector with approximately 10 pF/50 K

Ω load to GND; worst case

skew numbers are a calculation of worst case conditions through circuits. Both numbers apply to any channel
within a single or multiple module system.

[2] Channel 7 on the 3-state pods has been brought out in parallel as a non 3-state signal. By looping this output back

into the 3-state enable line, the channel can be used as a 3-state enable.

Agilent 10465A ECL Data Pod (unterminated)

Output type

10H115 (no termination)

Maximum clock

300 MHz

Skew [1]

typical < 1 ns; worst case = 2 ns

Recommended lead set

Agilent 10347A

Agilent 10466A 3-State TTL/3.3 volt Data Pod

Output type

74LVT244 with 100

Ω series; 10H125 on non 3-state channel 7 [2]

3-state enable

negative true, 100 K

Ω to GND, enabled on no connect

Maximum clock

200 MHz

Skew [1]

typical < 3 ns; worst case = 7 ns

Recommended lead set

Agilent 10474A

10H115

100

74LVT244