Clock pod characteristics – Agilent Technologies 16700 User Manual

Page 111

Advertising
background image

111

10463A ECL Clock Pod

Clock output type

10H116 differential unterminated; and differ-
ential with 330

Ω to –5.2V and 47 Ω series

Clock output rate

300 MHz maximum

Clock out delay

approximately 8 ns total in 14 steps (16720A
only); 11 ns maximum in 9 steps (16522A

only)

Clock input type

ECL – 10H116 with 50 K

Ω to –5.2v

Clock input rate

dc to 300 MHz

Pattern input type

ECL – 10H116 with 50 K

Ω (no connect is

logic 0)

Clock-in to clock-out

approximately 30 ns

Pattern-in to recognition

approximately 15 ns + 1 clk period

Recommended lead set

Agilent 10474A

10H116

–5.2 V

VBB

50 k

CLKin

10H116

–5.2 V

330

47

Ω CLKout

Pattern Generation Modules Specifications
and Characteristics

Clock Pod Characteristics

10460A TTL Clock Pod

Clock output type

10H125 with 47

Ω series; true & inverted

Clock output rate

100 MHz maximum

Clock out delay

approximately 8 ns total in 14 steps (16720A
only); 11 ns maximum in 9 steps (16522A

only)

Clock input type

TTL – 10H124

Clock input rate

dc to 100 MHz

Pattern input type

TTL – 10H124 (no connect is logic 1)

Clock-in to clock-out

approximately 30 ns

Pattern-in to recognition

approximately 15 ns + 1 clk period

Recommended lead set

Agilent 10474A

47

CLKout

10H125

WAIT

10H124

CLKin

Advertising