Individual data channel, Clock channel – Agilent Technologies 16700 User Manual

Page 96

Advertising
background image

96

State/Timing Modules Specifications and
Characteristics

Specifications for Each Input

Parameter

Minimum

Description/Notes

800, 1250, 1500 Mb/s modes

200, 400 Mb/s modes

Data

tWidth*

500 ps

1.25 ns

Eye width in system under test [2]

to Clock tSetup

250 ps

625 ps

Data setup time required before tSample

tHold

250 ps

625 ps

Data hold time required after tSample

All

vHeight [1]

100mV

100mV

E5379A 100-pin differential probe [3]

Inputs

250 mV

250 mV

E5378A 100-pin single-ended probe [4],
E5382A single-ended flying-lead probe set

300mV

300mV

E5380A 38-pin single-ended probe

User Adjustable Settings for Each Input

Parameter

Adjustment Range

1500 Mb/s mode

1250 Mb/s mode

800 Mb/s mode

400 Mb/s mode

200 Mb/s mode

Data

Adjustment Resolution

10 ps

10 ps

10 ps

100 ps

100 ps

to Clock

tSample [5]

0 to +4 ns

-2.5 to +2/5 ns

-2.5 to +2/5 ns

-3.2 to +3.2 ns

-3.5 to +3 ns

All

vThreshold [6]

10 mV resolution

10 mV resolution

10 mV resolution

10 mV resolution

10 mV resolution

Inputs

-3 to +5 V

-3 to +5 V

-3 to +5 V

-3 to +5 V

-3 to +5 V

Individual

Data Channel

vHeight

tWidth

tSetup

Sampling

Position

Clock Channel

Note (1)

tHold

vThreshold*

*User Adjustable

—0V—

tSample*

Data Eye

Figure 6.17. Data Sampling.

* All specifications noted by an asterisk in the table are the performance standards against which the product is tested.
[1] The analyzer can be configured to sample on the rising edge, the falling edge, or both edges of the clock. If both edges are used with a single ended clock input, take care to set the

clock threshold accurately to avoid phase error.

[2] Eye width and height are specified at the probe tip. Eye width as measured by eye finder in the analyzer may be less, and still sample reliably.
[3] For each side of a differential signal.

[4] The clock inputs in the E5378A and the E5382A may be connected differentially or single ended. Use the E5379A vHeight spec for clock channel(s) connected differentially.
[5] Sample positions are independently adjustable for each data channel input. A negative sample position causes the input to be synchronously sampled by that amount before each

active clock edge. A positive sample position causes the input to be synchronously sampled by that amount after each active clock edge. A sampling position of zero causes synchro-
nous sampling coincident with each active clock edge.

[6] Threshold applies to single-ended input signals. Thresholds are independently adjustable for the clock input of each pod and for each set of 16 data inputs for each pod. Threshold lim-

its apply to both the internal reference and to the external reference input on the E5378A.

vHeight

2X vHeight

vHeight

nSignal

pSignal

——0V——

Agilent Technologies 16760A
Supplemental Specifications* and Characteristics (continued)

Synchronous Data Sampling