Agilent Technologies 16700 User Manual

Page 102

Advertising
background image

102

Oscilloscope Modules Specifications and
Characteristics

16534A Characteristics*

Vertical (Voltage)

Vertical sensitivity range

16 mV full scale to 40 V full scale

Vertical resolution

8 bits full scale

Rise time (calculated from bandwidth)

700 ps

dc gain accuracy

±(1.25% of full scale + 0.08% per °C difference from
calibration temperature)

dc offset range
Vertical sensitivity

Offset range

• 16 mV full scale to 400 mV full scale

• ±2 V

• 400 mV full scale to 2.0 V full scale

• ±10 V

• 2.0 V full scale to 10 V full scale

• ±50 V

• 10 V full scale to 40 V full scale

• ±250 V

Probe attenuation

Any ratio from 1:1E-9 to 1:1E+6

Channel-to-channel isolation (with channel sensitivities equal)
• dc–50 MHz

• 40 dB

• 50 MHz–500 MHz

• 30 dB

Maximum safe input voltage
• 1 M

• ±250 V dc + peak ac (<10 kHz)

• 50

• 5 Vrms

* Characteristics refer to the input at the BNC connector