Agilent Technologies 16700 User Manual

Page 87

Advertising
background image

87

State/Timing Modules Specifications and
Characteristics

Agilent Technologies 16710A, 16711A, 16712A
Supplemental Specifications* and Characteristics

Probes (general-purpose lead set)

Input resistance

100 K

Ω, ±2%

Parasitic tip capacitance

1.5 pF

Minimum voltage swing

500 mV, peak-to-peak

Threshold accuracy*

±(100 mV + 3% of threshold setting)

Maximum input voltage

±40 V peak

State Analysis

Minimum state clock pulse width

3.5 ns

Time tag resolution [1]

8 ns

Maximum time count between states

34 seconds

Maximum state tag

4.29 x 109 states

count between states [1]

Minimum master to master clock time*

16710A, 16711A, 16712A: 10 ns

Minimum master to slave clock time

0.0 ns

Minimum slave to master clock time

4.0 ns

Context store block sizes

16, 32, 64 states

16710A/11A/12A only

Timing Analysis

Sample period accuracy

0.01% of sample period

Channel-to-channel skew

2 ns, typical

Time interval accuracy

± (sample period + channel-to-channel
skew + 0.01% of time interval reading)

Minimum detectable glitch

3.5 ns

* All specifications noted by an asterisk are the performance standards against which the product is tested.
[1] Time or state tags halve the acquisition memory when there are no unassigned pods.

1.5pF

370 ohms

7.4pF

GROUND

100K
ohm

Figure 6.12. Equivalent probe load for the
Agilent 16710A, 16711A and 16712A, general-
purpose lead set.

Advertising