KYORITSU 3128 User Manual

Page 16

Advertising
background image

12

KEW3128

3. Specification

KEW3128

Volt Meter

Range

Voltage Test

Measuring Range

DC Voltage

AC Voltage

±30~±600V

30~600V(50/60Hz)

Accuracy

±2%rdg±3dgt

Frequency

Range

Voltage Test

Measuring Range

45.0 ~ 65.0Hz

Accuracy

±0.2Hz

Ammeter

Measuring Range

0.0nA ~ 2.40mA

(determined by

resistance and voltage

values)

Depending on the

effective range of

insulation resistances

Capacity Meter

Range

500V ~ 5000V Range

10000V / 12000V Range

Accuracy

5.0nF ~ 50.0μF

5.0nF ~ 1.0μF

±5%rdg±5dgt

* Measurements of capacitance of 0.5μ or more shouldn’t be repeated at

10000V/ 12000V Ranges in a short time. (Rough guide: 5 times/ hour)

* A message “Noise Error” may be displayed on the LCD and a measurement

may be ceased at 10000/12000V Ranges. In this case, select a lower Voltage

Range and test again.

Calculated Value
PI,DAR,DD

Measurement Mode

PI

DAR

DD

Display Range

0.00 ~ 999

0.00 ~ 999

0.00 ~ 999

Computational error

±2dgt

±2dgt

±2dgt

 ●

Electromagnetic compatibility (IEC61000-4-3)

  

Radio-frequency electromagnetic field

= 10V/m : 20times of the specified accuracy

Advertising