9 capacitance measurement 6.9.1 measurement screen – KYORITSU 3128 User Manual

Page 71

Advertising
background image

KEW3128

67

KEW3128

6. Measurement

6. 9 Capacitance Measurement

6.9.1Measurement Screen

Displayed Items

Details

Capacitance value

Displays capacitance values of the measured
object after insulation resistance tests.

Measuring Time

Elapsed time after a start of measurement

  

At capacitance measurements, measured values are displayed when
insulation resistance measurements complete. When the output voltages
are 80% or less of the preset voltage values at an insulation resistance
measurement, readings for capacitance become“---”.

  

KEW3128 has a Protect Mode to limit charge currents in order to protect

the instrument when measuring 10uF or higher. In this mode, a message

Protect mode”is displayed on the LCD.

  

The instrument exits from the Protect Mode automatically when a battery
charge completes or 5 min pass after entering into this mode.