KEW3128

7. Battery Charging and Replacement

7. Battery Charging and Replacement

7. 1 How to charge battery

#

DANGER

Use the special cord supplied with this instrument only. Firmly connect the

Power Cord to an outlet. Never connect it to a device on which electrical
potentials higher than AC240V exist.
Handling and storage instructions specified by the battery manufacturer
should be observed.

#

WARNING

Connect the Power Cord to the instrument first. Cord to be firmly inserted.
Do not use the Cord if any abnormal conditions such as cracks or exposed
metal parts are present. When unplugging the Cord from the mains socket
outlet, do so by removing the Plug first and not by pulling the Cord.

  ① 

Set the Range Switch to the OFF position.

  ② 

Confirm a battery is installed in the instrument.

  ③ 

Connect the Power Cord to the instrument to feed power to the
instrument.

  ④ 

LED Status Indicator flashes in red and the Battery Mark also flashes
on the LCD.

  ⑤ 

The indicator lights up in green and the Battery Mark on the LCD stops
flashing and lights up. (Battery charge completes in about 8 hours.)

* Battery life and how many times can be charged are depended on conditions of

use and environment.

* Storing rechargeable lead-acid batteries in a low-charged state could lead to

reduced life and/or damage. When storing the battery for a long period, check
and charge the battery at regular intervals.