Measurement time, Condition current consumption – KYORITSU 3128 User Manual

Page 2

Advertising
background image


P.11 Additional graph of

“Resistance-Output voltage Characteristic” for each Voltage range

[Graph of Resistance-Output voltage Characteristic]






Rated current at 500V, 1000V, 2500V, 5000V Ranges : max 1mA or higher

Rated current at 10kV, 12kV Ranges : max 0.5mA

Above graphs show the relations between output voltages and measurement resistances.

*

“Rated measurement current” described in the table at page 11 should be corrected to “Output current”.

P.15 Additional description about current consumption and battery life

 Continuous measurement: No limitation (IR measurement mode)

* recorded data and graph are max 90 min.

Max 90 min (SV measurement mode)

Max 60 min (PI/DAR/DD measurement mode)

 Max current consumption and measurement time while rated currents are maintained.

Condition

Current consumption

Measurement time

500V /

300kΩ

2100mA

or less

Approx

2 hours

1000V / 600kΩ
2500V / 2.4MΩ
5000V / 4.8MΩ
10000V / 20MΩ
12000V / 24MΩ

* Current consumption may exceed above values when low resistances, which influence the output of

rated currents, are measured.

* Measurement time (approx 4 hours) described at page 15 and the Measurement time in above table

are the periods that full-charged battery voltage drops to the lowest level.

* It is recommended to charge the battery with reference to

“7.1 How to charge battery” described in the

manual before stating to use with the instrument since the battery voltage may be low due to

self-discharge.


10kV/12kV Range

0

2,000

4,000

6,000

8,000

10,000

12,000

14,000

0.0001

0.001

0.01

0.1

1

10

100

Insulation Resistance (G

Ω )

O

u

tp

u

t

Vo

lt

ag

e

(V)

12kV

10kV

500V/1000V/2500V/5000V Range

0

1,000

2,000

3,000

4,000

5,000

6,000

0.0001

0.001

0.01

0.1

1

10

100

Insulation Resistance (G

Ω )

O

u

tp

ut

V

olt

ag

e

(V

)

500V

1000V

2500V

5000V

Advertising