6 filter mode, 7 save data, Types of save data – KYORITSU 3128 User Manual

Page 45

Advertising
background image

KEW3128

41

KEW3128

6. Measurement

6.1.6 Filter Mode

The KEW3128 has following 3 kinds of Filter Functions.

Filter Mode is effective to reduce the variations in readings due to external

influences during high resistance measurements.

Effectiveness of the Filter Mode becomes stronger when values get bigger. To

check sudden variations in resistances, the Filter Mode should be disabled.

Name

Function

Filter OFF

Disables the Filter (default setting)

Filter 1

Low-pass Filter (fc = 0.3Hz)

Filter 2

Moving Average (average of 5 data)

Filter 3

Low-pass Filter + Moving Average

6.1.7Save Data

  

Types of Save Data

   

The KEW3128 handles following 3 types of data.

  ●

Logging Data (REC File)

   

Record the measured values (voltage, current and resistance values) at

every 1 sec from the beginning to the end of a measurement.

   

The max recording period is 90 min. The “   ” Measurement

Recoding should be selected on the ENTER Menu (=>P.37) while the

instrument is in the Stand-by mode to save the Logging data.

   

The time displayed at the top of the LCD indicating the time left that data

can be recorded. (=>P.64)

   

Data is saved as “RECXX”. (XX : 01~32)

  ●

Measured Data (SAVE File)

   

Measured data contains measured results only.

   

Select the

Saving Measured Results on the ENTER Menu

(=>P.37) while the measured results are displayed on the LCD.

   

Data is saved as “SAVEXX”. (XX : 01~32)

  ●

Print Screen (BMP File)

   

Captures and saves the screen images. A long press (1 sec or longer)

of the Print Screen/ Backlight Key saves screen images.

   

Data is saved as “BMPXX”. (XX : 01~32)

Advertising