Section 3 - Electrical Connections and Setup

Page 8

MULTI-PARAMETER TRANSMITTER USER’S MANUAL

Section 3 - Electrical Connections and Setup

3.4 pH or ORP Combination Probe connections and setup

The drawing shows the connections for the Aquametrix Combination probe. The cable should be run
in a conduit separate from AC power wires, and via a separate conduit hole. The cable length should
not exceed 10 feet (3 meters).

Fig. 3.3 Connections for the 2 and 4 wire Combination Probe

SEC. 5.8

SEC. 5.5

SEC. 5.1

ORP PROBE

SELECT COMBINATION PROBE

SELECT ORP METER

MANUAL CALIBRATE

PH PROBE

MANUAL CALIBRATE

SELECT PH METER

SELECT COMBINATION PROBE

SEC. 4.1

SEC. 4.6

SEC. 4.7

SEC 4.10

MANUAL PROBE
TEMPERATURE SETUP

DISABLE TEMPERATURE

SEC 5.10

DISPLAY

MANUAL CAL PH

METER SELECTION

PROBE SELECT

7.15pH 25.0C

T.COMP OVERRIDE

7.15pH 25.0C

PROBE SELECT

METER SELECTION

MANUAL CAL PH

T.DISP OVERRIDE

RUN MODE

TWO WIRE PROBE WITH NO TEMP SENSOR

TWO WIRE PROBE WITH NO TEMP SENSOR

RUN MODE

ORP

pH

Once connected, step through the LCD menus to select the probe in the order shown. The first two steps
may be skipped if the meter is already configured for a Combination Probe. If a two wire pH probe is
used, which has no temperature sensor, ensure that the Temp. Comp. Override is set to same tempera-
ture as the buffer before calibrating. If a two wire ORP probe is used, you can blank the Temp display
with the T DISP OVERRIDE menu.

The

2 wire version has

no temperature sensor
and is connected via a
coaxial wire.
In a pH meter, the user
should set the

T COMP

OVERRIDE menu to ON
(Section 4.10) and adjust
the temperature setting to
the actual probe temper-
ature.
In an ORP meter, the
user should set the
T.DISP OVERRIDE to
ON (Section 5.9) to blank
the temperature reading
on the display.

The

4 wire version has

two additional wires for
the probe internal tem-
perature sensor. Ensure
that the

T COMP OVER-

RIDE or T.DISP OVER-
RIDE
is OFF.

Dwg# N104-43