Ph - calibration menu - auto calibrate 4.2, Multi-parameter transmitter user’s manual, Ph - calibration menu - auto calibrate 4.2 page 15 – Metex SharkTX Multi-Parameter Transmitters User Manual

Page 17

Advertising
background image

MULTI-PARAMETER TRANSMITTER USER’S MANUAL

pH - Calibration Menu - Auto Calibrate 4.2

pH - Calibration Menu - Auto Calibrate 4.2

Page 15

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

If an erros occurs, the controller will indicate a
"BUFFER ERROR" alarm. This could be caused by
- using a defective buffer
- incorrect probe wiring
- defective probe

Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the
diagnotics menus to test the probe
see Section 4.16

If an erros occurs, the controller will indicate a
"BUFFER ERROR" alarm. This could be caused by
- using the same buffer twice
- using a defective buffer
- incorrect probe wiring
- defective probe

Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the
diagnotics menus to test the probe

If the calibration did not appear to
be correct, press the key
which will return back to the Auto cal
menu.

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

This will store the calibration

data and return back to the

Auto cal setup so the user

can select another function.

If the calibration is OK, use the key to

move the cursor over the Y text and press the

down key.

After 5 seconds, the controller will

compute the slope of the calibration, the

estimated probe efficiency and the probe

temperature.

When complete, the controller will report the

measured value

When complete, the controller will report

the measured value

Press the key to move to Buffer 2

Please wait for the controller to complete

the measurement

The controller will read the pH value,

averaging a number of results to get a

stable calibration value.

Place the probe in the second buffer

solution.Be sure to clean and rinse the

Probe first with D.I. water and then insert

it in the 4.00 buffer.

Press the key

Please wait for the controller to complete

the measurement

The controller will read the pH value,

averaging a number of results to get a

stable calibration value.

Place the probe in the first buffer solution,

be sure to clean and rinse the Probe first

with D.I. water and then insert it in the 7.00

buffer.
Press

RUN MODE

AUTO CAL pH

STORE? Y N

STORE? Y N

AUTO CAL pH

AUTO CAL pH

EFF 95% 24.8C

SLOPE 61.22MV/pH

BUFFER2 4.00 >

AUTO CAL pH

AUTO CAL pH

BUFFER ERROR

CONTINUE >

RUNNING AUTO CAL
BUFFER2 WAIT...

BUFFER1 7.00 >

PRESS 'DOWN'

IF BUFFER2 READY

BUFFER ERROR

CONTINUE >

PRESS 'DOWN'

IF BUFFER1 READY

RUNNING AUTO CAL
BUFFER1 WAIT...

CALIBRATION

7.15pH 25.0C

MANUAL CAL pH

This example shows an AUTO
calibration with buffers of 7pH & 4pH.

When performing an auto calibration,
any two standard buffers of 4.00, 7.00
or 10.00 pH can be used. The buffers
will be automatically recognized if the
probe measures them within +/- 1 pH
of their nominal value.

Advertising
This manual is related to the following products: