ORP - Utilities Menu - Manual Test Mode 5.3

Page 33

MULTI-PARAMETER TRANSMITTER USER’S MANUAL

ORP - Utilities Menu - Manual Test Mode 5.3

DOWN

DOWN

DOWN

UP

DOWN

DOWN

UP

DOWN

Manual Test Mode is used to simulate a process reading
in order to verify the correct response of the output.
When in the Manual Test Mode, the outputs is no longer
placed on hold as it is when in the rest of the menu.

The range is 0 to +1000mV.

NOTE:

When the user exits the Manual Test Mode, the 4-20mA

output will remain in the Test Mode state until the operator

enters the run mode. The output will then revert back to

the previous On-line state.

The 4-20 mA output will track
the change in process value

RUN MODE

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

Press the key once to
move the cursor to the RH side
of the display

The 4-20 mA output will follow the process
value change.

Press the key to return to the
MANUAL TEST MODE menu

Use the and keys to change
the simulated process value.

Press the key once which will
move the cursor over the least digit of the
simulated process value.

TEST 50 0 >