Conductivity - Calibration Menu - Temp. Calibration 6.3

Page 49

MULTI-PARAMETER TRANSMITTER USER’S MANUAL

Conductivity - Calibration Menu - Temp. Calibration 6.3

DOWN

DOWN

DOWN

DOWN

UP

DOWN

DOWN

DOWN

UP

DOWN

DOWN

DOWN

If the Temperature Compensation Override is set to ON
(see section 6.11), the Shark cannot calibrate the
temperature sensor. This display will appear to alert the
user to the condition.

If the calibration did not appear

to be correct, press the

key which will return back to the

Temp Calibration menu.

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

This will store the calibration

data and return back to the

Temp Calibration setup so

the user can select another

function.

Use the and to adjust

the reading until it agrees with the

known temperature of the probe

Press the key to accept the

reading and move to the store function

"T FACTOR SHIFTED" will be

displayed for 5 seconds to indicate that

the temperature sensor has been

calibrated.

The Shark will then move to the store

function.

The controller displays the current probe

temperature.

If this incorrect, the controller can be

adjusted to compensate.

Press the key once

which will move the cursor

back to the RH side of

temperature display.

Press the key once

which will move the cursor

over the least significant digit

of the temperature display.

RUN MODE

DRY CAL COND

STORE? Y N

TEMP CALIBRATION

STORE? Y N

TEMP CALIBRATION

TEMP CALIBRATION
25.0C >

T FACTOR SHIFTED

25. 0 C >

TEMP CALIBRATION

TEMP CALIBRATION

TEMP CALIBRATION

TEMP CALIBRATION

22. 4 C >

22.4C >

TEMP CALIBRATION
TEMP O/R ON >

1000uS 25.0C

CALIBRATION

MANUAL CAL COND

DOWN

DOWN

Temperature Calibration
In most cases, the factory temperature calibration is accurate enough to ensure correct temperature readings. However, in some circumstances, the user
may wish to ensure the temperature sensor is calibrated accurately, especially when operating at the extreme end of the conductivity cell temperature
operating range, or where the temperature compensation is critical to correct process readings.
This menu allows the user to calibrate the temperature anywhere within it's range.
Be aware, that the conductivity reading is affected by the temperature reading (due to the temperature compensation) so accurate temperature
calibration is vital to obtaining accurate conductivity readings. If the user is unsure of the calibration test fixture, then it would be best to leave
the temperature calibration at it's factory setting.
Be sure to allow the temperature of the cell to stabilize before attempting to calibrate the temperature sensor, this may take a significant amount of time as
the sensor is buried behind a protective layer of epoxy which will cause some delay.