Conductivity - Calibration Menu - Manual Calibrate 6.1

Page 47

MULTI-PARAMETER TRANSMITTER USER’S MANUAL

Conductivity - Calibration Menu - Manual Calibrate 6.1

DOWN

DOWN

DOWN

UP

DOWN

Manual Calibration is used to "wet calibrate the cell". This can be done with the cell installed in the
process, or with the cell suspended in a known buffer solution.
When calibrated "In Process", the actual conductivity is determined with a grab sample or a hand held
meter, and the value entered in the display.
When calibrated with buffers, the cell is placed in a known buffer solution, and the value of the buffer
entered on the display.
In both cases, make sure the cell has time to stabilize both in temperature and conductivity before
entering any data.

If the calibration did not appear
to be correct, press the
key which will return back to the
Manual Cal Cond menu.

DOWN

RUN MODE

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

Place the cell in the buffer solution

Press to store the
calibration data and return
back to the Manual Cal Cond
menu so the user can select
another function.

The controller will read the
Conductivity value, averaging a
number of results to get a stable
calibration value.

Use the and keys to
adjust the reading until it agrees with
the actual buffer conductivity
value

If the calibration is OK, use the
key to move the cursor over the Y
text and press the down key.

Then press to move
the cursor to the RH position

Press the key once
which will move the cursor
over the least sign. digit of
the display.

UP

DOWN

DOWN

CALIBRATION

1000uS 25.0C

MANUAL CAL COND

STORE? Y N

STORE? Y N

MANUAL CAL COND

MANUAL CAL COND
100 0 uS

>

1000 uS

>

MANUAL CAL COND

MANUAL CAL COND

967 uS

>

MANUAL CAL COND

96 7 uS

>

MANUAL CAL COND