Orp - calibration menu - manual calibrate 5.1, Multi-parameter transmitter user’s manual – Metex SharkTX Multi-Parameter Transmitters User Manual

Page 33

Advertising
background image

ORP - Calibration Menu - Manual Calibrate 5.1

Page 31

MULTI-PARAMETER TRANSMITTER USER’S MANUAL

ORP - Calibration Menu - Manual Calibrate 5.1

DOWN

UP

DOWN

DOWN

DOWN

DOWN

UP

DOWN

DOWN

DOWN

DOWN

DOWN

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

Press to store the
calibration data and return
back to the Manual Cal menu
so the user can select another
function.

Place the probe in the buffer solution, be sure
to clean and rinse the Probe first with D.I.
water and then insert it in the mV buffer.

Press

If the calibration did not appear
to be correct, press the
key which will return back to the
Manual Cal menu.

The controller will display the mV value and the
probe temperature for 5 seconds.
If the T.DISPLAY O/R is enabled, the display will
show.

If the calibration is OK, use the
key to move the cursor over the Y
text and press the down key.

Then press to move
the cursor to the RH position

Use the and keys to
adjust the reading until it agrees with
the actual buffer mV value

When complete, the controller will
report the measured value

Please wait for the controller to
complete the measurement

The controller will read the mV value, averaging a
number of results to get a stable calibration value.

RUN MODE

IF BUFFER READY

CALIBRATION

MANUAL CAL ORP

STORE? Y N

STORE? Y N

TEMP 25.0C

BUFFER 600mV

MANUAL CAL ORP
BUFFER 600

>

RUNNING MANU CAL
BUFFER WAIT...

PRESS ©DOWN©

BUFFER 50 0

>

MANUAL CAL ORP
BUFFER 60 0

>

MANUAL CAL ORP

MANUAL CAL ORP

TEMP O/R ON

MANUAL CAL ORP

500mV 25.0C