Setra System SRCM User Manual

Page 10

Advertising
background image

10

By pressing the touch-screen directly on the word POSITIVE (or NEGATIVE,

NEUTRAL, or STANDBY), the user is able to change the condition of the

room between ACTIVE and STANDBY modes. A pop-up menu appears to en-

able selection. If passwords are enabled, the user must first enter a password to

proceed with the change.

In ACTIVE mode, the full function of the SRCM is active. In STANDBY

mode, the SRCM will perform all functions except that alarming will be dis-

abled (both audible and visual alarms). The STANDBY mode is designed to

put the SRCM into a mode where room pressurization is not critical, such as

cleaning, patient transfer, or longer-term unoccupied status. Note that this op-

eration can be password protected so that only an Operator or a Supervisor can

make this change. Passwords are set up in Setup Display Advanced tab.

Active and Standby
Modes

Advertising