Agilent Technologies E8247C PSG CW User Manual

Page 102

Advertising
background image

92

Chapter 5

Dual Arbitrary Waveform Generator
Arbitrary (ARB) Waveform File Headers

3. Return to the ARB Setup menu: Press

Return

.

This menu lets you change the current instrument settings.

Figure 5-3

shows the ARB Setup softkey

menu and the softkey paths used in steps four through nine.

4. Set the ARB sample clock to 5 MHz: Press

ARB Sample Clock

>

5

>

MHz

.

5. Set the modulator attenuation to 15 dB:

Press

More (1 of 2)

>

Modulator Atten n.nn dB Manual Auto

to Manual >

15

>

dB

.

6. Set the I/Q modulation filter to a through:

Press

I/Q Mod Filter Manual Auto

to Manual >

Through

.

7. Set marker one to blank the RF output at the set marker point(s):

Press

More (2 of 2)

>

Marker Utilities

>

Marker Routing

>

Pulse/RF Blank

>

Marker 1

.

For information on setting markers, see

“Using Waveform Markers” on page 104

.

8. Set the polarity of Marker 1 negative: Press

Return

>

Marker Polarity

>

Marker 1 Polarity Neg Pos

to Neg.

9. Return to the Header Utilities menu: Press

Return

>

Return

>

Header Utilities

.

Notice that the

Current Inst. Settings

column now reflects the changes made to the current

signal generator setup in steps 4 through 8, but that the saved header values have not changed (as shown
in

Figure 5-4 on page 94

).

10. Save the current settings to the file header: Press

Save Setup To Header

softkey.

The settings from the

Current Inst. Settings

column now appear in the

Saved Header

Settings

column. The file header has been modified and the current instrument settings saved. This is

shown in

Figure 5-5 on page 94

.

While a modulation format is active, the waveform file (AUTOGEN_WAVEFORM) is playing and you can
modify the header information within the active modulation format. Once you turn the modulation format
off, the header information is available only in the dual ARB player. If you turn the modulation format off
and then on, you overwrite the previous AUTOGEN_WAVEFORM file and its file header. To avoid this,
rename the file before you turn the modulation format back on (see

page 103

).

When you store the waveform file (see

page 103

), the header information is stored with the waveform.