Chapter 10

191

Troubleshooting

RF Output Power Problems

Figure 10-2

Reverse Power Solution

Compared to the original configuration, the ALC level is 10 dB higher while the attenuator reduces the LO
feedthrough (and the RF output of the signal generator) by 10 dB. Using the attenuated configuration, the
detector is exposed to a +2 dBm desired signal versus the
-15 dBm undesired LO feedthrough. This 17 dB difference between desired and undesired energy results in
a maximum 0.1 dB shift in the signal generator’s RF output level.

RF LEVEL

CONTROL

SIGNAL GENERATOR
OUTPUT CONTROL

MIXER

LO

IF

ALC LEVEL/

DETECTOR
MEASURES
+2 dBm
ALC LEVEL

DETECTOR
MEASURES
- 15 dBm
REVERSE
POWER

LO FEEDTHRU
= - 5 dBm

RF INPUT
= - 8 dBm

LO LEVEL
= +10 dBm

10 dB

ATTEN

RF OUTPUT
= +2 dBm

Advertising