Agilent Technologies E8247C PSG CW User Manual

Page 49

Advertising
background image

Chapter 2

39

Basic Operation

Configuring the RF Output

6. Preset either instrument.

Presetting one of the instruments should automatically preset the other as well. If both instruments do
not preset, check the GPIB connection, GPIB addresses, and ensure the 8757D is set to system interface
mode (

SYSINTF

set to ON).

The PSG automatically activates a 2 GHz to maximum frequency ramp sweep with a constant amplitude
of 0 dBm. Notice that the

RF ON

,

SWEEP

, and

PULSE

annunciators appear on the PSG display. The

PULSE

annunciator appears because the 8757D is operating in AC mode.

The PSG also switches its remote language setting to 8757D System, allowing the PSG to talk to the
8757D during ramp sweep operations. You can confirm this by pressing

Utility

>

GPIB/RS-232 LAN

and

observing the selection under the

Remote Language

softkey.

NOTE

During swept RF output, the

FREQUENCY

and/or

AMPLITUDE

areas of the signal

generator’s display are deactivated, depending on what is being swept. In this case,
since frequency is being swept, nothing appears in the

FREQUENCY

area of the display.

7. Press

Frequency

>

Freq CW

.

The current continuous wave frequency setting now controls the RF output and ramp sweep is turned off.

8. Press

Freq Start

.

The ramp sweep settings once again control the RF output and the CW mode is turned off. Pressing any
one of the softkeys

Freq Start

,

Freq Stop

,

Freq Center

, or

Freq Span

activates a ramp sweep with the current

settings.

NOTE

In a frequency ramp sweep, the start frequency must be lower than the stop frequency.

9. Adjust the settings for

Freq Center

and

Freq Span

so that the frequency response of the device under test

(DUT) is clearly seen on the 8757D display.

Notice how adjusting these settings also changes the settings for the

Freq Start

and

Freq Stop

softkeys. You

may need to rescale the response on the 8757D for a more accurate evaluation of the amplitude.

Figure

2-3 on page 40

shows an example of a bandpass filter response.